Custom Essays and Free Coursework

The UK's Favourite Provider of Custom Essays, Custom Dissertations, Free Coursework, Model Answers, University Assignments.

degree essays logo

Research Brief 4234 – Mechanical Breathing Machines

1. The development of the breathing machine

The breathing machine – or artificial ventilator – is described by Young and Sykes (1994) as a mechanical device that replaces the function of the inspiratory (breathing-in) muscles. In reality it is far more sophisticated than the description alludes to. It has a primary use in medical care and in particular for patients with breathing difficulties, respiratory complaints and those that require anaesthetising for operation.

The history of air resuscitation differs according to which author one chooses to believe but the earliest recording writes Pilbeam (1986) is from the early 16th century when Paracelcus used a fire bellows in connection to a tube inserted into the patients’ mouth. It is unclear as to who is regarded as the inventor of the ventilator not requiring a human power source. Goerig et al (1987) credit George Fell with the invention of the artificial breathing machine. Rendell-Baker and Pettis (1987) then write that Draeger developed the first commercially available artificial ventilators. Experimentation continued until a polio epidemic in Copenhagen in 1952 in which 1400 university students were employed in shifts to manually operate these artificial ventilators. This scare triggered a great surge in research for an automatic breathing machine.

In 1967 the first mechanical and automatic positive pressure ventilation machine was introduced. Positive pressure acts to increase the pressure in the airways. Negative pressure ventilators reduce pressure around the thorax using rigid chambers that enclose the thorax or the whole body neck down. These are usually known as tank respirators or “iron-lungs”. They are outdated in modern times though simply due to their size and the lack of access for the patient. Their use is minimal in contemporary healthcare and for this reason the remainder of this piece will focus upon positive pressure ventilators.

2. Biological theory and requirements

Although the biological theory behind artificial ventilators is complex the following section will show a simplified process of human breathing. According to Kestner (1981), inspiration is caused by a pressure gradient between the atmosphere and the alveoli due to contraction of the inspiration muscles. This pressure gradient generates airflow into the lung until the point when the alveolar pressure is equal to atmospheric pressure. Expiration (breathing-out) is a passive process and as the muscles relax an opposite pressure gradient is created which causes an airflow to leave the lungs. This is known as spontaneous breathing. But what happens when this process cannot take place?

What happens is that a mechanical ventilator is used to, in effect, do the breathing for the patient. Al-Shaikh and Stacey (2001) define properties of the ideal breathing system as;

· Simple and safe to use.
· Delivers the intended inspired gas mixture.
· Permits spontaneous, manual and controlled ventilation in all age groups.
· Efficient, requiring low fresh gas flow rates.
· Sturdy, compact and lightweight in design.
· Permits the easy removal of waste exhaled gases.
· Easy to maintain with minimal running costs.

So, let us now move on to investigate the different types of ventilator that there are currently in use.

3. Ventilation Classifications

3.1 Standard features

Most ventilators use standard components in reservoir air bags, adjustable pressure limiting (APL) valves, and tubing. The APL valve is a one-way valve that allows exhaled air and excess fresh gas to leave whilst disallowing the entry of atmospheric gases. The valve opens and releases gas when it reaches a certain pressure and this defined pressure can be altered. The main task of the reservoir bag is to accommodate fresh gas intake and to limit pressure build up within the system. The use of tubing should be straightforward.

3.2 Classification
The most common classification used to differentiate between systems using the previous components is the Mapleson (1962) classification. This classification is based upon two characteristics;

a. The method by which gas is driven into the lung – flow or pressure
b. The cycle between inspiration and expiration – timed or thresholded

The following diagram shows the different classifications of ventilatory modes. It should be noted that these are not breathing machines per se – they are the physical modes of artificial breathing.


Figure 1. Classification of Positive Pressure Ventilatory Modes
(Aitkenhead and Smith, 1996)


Note: The arrow indicates fresh gas supply

Referring to the above diagram, according to Al-Shaikh and Stacey (2001) modes B and C are used for recovery and emergency whilst modes A,D and E are used for anaesthesia. Mode A, also known as the Magill system, works by the following method; As the patient exhales the gas is pushed through the tubing towards the reservoir bag which is filled continuously with a flow of fresh gas. The pressure build up causes the APL valve to open and expel the alveolar gas. At this point the patient should have inspirated once more thus receiving a mixture of the fresh gas and rebreathed gas. Mode D is similar but utilises a coaxial tube to move breathed and fresh gas without mixing them. Modes B and C operate similarly but uses greater use of fresh gas input to avoid rebreathing which is important for recovery situations. Modes E and F are used primarily in paediatrics as resistance to expiration of the mode is minimal which is obviously important when treating this group.

4. The Mechanics and Physics
4.1 Power

The previous section detailed a simplified view of ventilatory modes and typical components of the ventilation loop of a breathing machine. Actual mechanical breathing machines are more complex than this as they incorporate power and feedback mechanisms. Most ventilators are either electrically or compressed-gas powered which in turn drives the mechanism designed to ventilate. A typical drive mechanism will consist of a wheel, rod and piston – which delivers the gas.

Whilst the gas power transmission is fairly straightforward modern-day ventilators are controlled by micro-processor servos. There are essentially three types of control according to Branson et al(1995) which are pressure, volume or flow controllers. In order to fully understand the mechanics of an assisted breathing process we must understand the biology and physics. The process is based around the need to achieve the correct pressure to cause a flow of gas to increase the volume of the lungs.

4.2 Physical variables
Pressure, flow and volume are all variables in the process. For example and in direct relation to the problem in hand, the flow rate can be kept constant. A constant volume flow rate is generated by the piston – by altering the stroke length - and delivered to the lung. However, there is a problem. There are two factors that affect the flow rate, and therefore the pressure and volume. Firstly, resistance encountered as the gas enters the throat – the airways are resistant in terms of gas density and viscosity to name but a few. Secondly, something called compliance. The relationship between the gas volume and the pressure of alveolar gas is important. It depends upon the elastic properties of the alveoli and is described by compliance. Compliance is, writes Pilbeam (1986), the change in volume corresponding to the change in pressure accompanying the volume change. In relation to the desired constant volume flow rate we will no longer have this constant flow due to the effects of resistance and compliance. Whilst a constant flow rate is generated-known as tidal flow - it is not delivered. This pressure variance in the lungs is dependent upon the volume flow rate. Something must be done to compensate for this change and ensure that the flow rate generated is also delivered.

4.3 Schematics
There are many machines that can do this on the market today – such as Draeger . Using a flow controller, when resistance and compliance change, pressure will change but the volume will not. Piston ventilators compensate for this compliance by measuring it and delivering an appropriately proportional additional volume of flow to ensure that the desired tidal flow is delivered to the lung. The diagram below is a typical schematic;


Figure 2. Schematic of Piston Ventilation Circuit
Adapted from Draeger (2003)

Using the schematic as a guide and taking the ventilatory pressure of the human to be zero – ie. the patient cannot breathe spontaneously – it can be seen how the ventilator functions. Fresh gas is sent into the system and the piston acts as the force upon this gas to push it through the inspiratory valve, which is calibrated to a set level, and into the lungs. As pressure reaches a certain level the gas is expired and forces its way through the expiration valve. Some of the gas is absorbed at this point and the cycle begins again. It should be noted that in order to compensate for compliance and resistance the desired measurement is taken as seen on the diagram. This value is then used to automatically calculate the desired additional volume, for example, of fresh gas required to maintain the tidal volume flow from generation to delivery. The standardised process is shown in the diagram below;




Figure 3. Block diagram of control system
Branson et al (1995)


The control system may be mechanical, pneumatic, fluidic, electric or electronic. Its basic function is to use sensors to measure the error signal – caused by resistance and compliance – and to automatically feedback and adjust the input to achieve the desired output. As mentioned before the control variable may be either pressure, volume or flow. Time is sometimes used also .

5. Constant flow machines
One example of a constant flow controller on the market is the Siemens Servo 900C (Branson et al 1995; Young and Sykes 1994; Ingelstedt et al 1972). As the name implies it uses servo control in measuring flow in order to adjust the output control valve accordingly to maintain a constant flow of air as the load – compliance and resistance – change. This particular machine uses two of both pressure and flow transducers – one at inspiration and one at expiration. The flow transducers are electronically integrated to measure volume. These volume signals can read both inspired and expired tidal volume and feedback to the main control point at which the inspiration flow is altered in line with what is desired in compensation. For example if constant flow is required the inspiratory valve is opened until the desired flow is reached. If the sensors find that this flow is decreasing due to the effects outlined then the valve will automatically be opened further to compensate and allow increased gas flow.

Alternatively to the schematic (figure 2) this particular machine is powered by an internal compressor in the form of bellows that is pressurised by a spring. The spring force sets the working pressure of the ventilator.

6. Literary Summary
Breathing machines can be traced back hundreds of years from the simple use of bellows right through to modern day advancements such as the positive pressure automatic ventilator or even further to the soda-lime ventilator described by Al-Shaikh and Stacey (2001) as using soda-lime to absorb carbon dioxide and so ensure the system uses as fresh gas as possible.

They are used to assist, control and maintain artificial breathing where the human may either be unable to breathe properly, or indeed, at all. The main applications are in anaesthesia and for recovery and emergency situations.

Most machines are powered by either a piston or bellows which is used to act as the force upon a separate supply of fresh gas. This new gas is inspirated by the patient via inward tubing and consequently expired via outward tubing.

Modern machines use complex electronic feedback and control systems in order to regulate and maintain proper use of ventilation. The process is complicated by the interference of resistance and alveolar compliance but usage of sensors can measure and feedback in order to correct the error and maintain a constant output – be it pressure, volume or flow – by valve adjustment.

  • REFERENCES

  • Aitkenhead, R., Smith, G., (1996), Textbook of Anaesthesia 3rd Edition, Churchill Livingstone
  • Al-Shaikh, B., Stacey, S. (2001) Essentials of Anaesthetic Equipment 2nd Edition, Harcourt Publishing
  • Branson, R.D., Hess, D.R., Chatburn, R.L., (1995) Respiratory Care Equipment, JB Lippincott Company Publishing
  • Draeger (2003), Series on Advanced Ventilation, [online] at URL: http://www.draeger.com/us/MT/Library/or/case_studies/advanced_vent.pdf, last viewed 8/5/03
  • Goerig, M., Filos, K., Ayisi, K.W., (1987), George Edward Fell and the Development of Respiratory Machines, In: Atkinson, R.S., Boulton, T.B., (Eds.), (1987), The History of Anaesthsia, Royal Society of Medicine
  • Ingelstedt, S., Jonson, B., Nordstrom, L., Olsson, S.G., (1972) A Servo-Controlled Ventilator Measuring Expired Minute Volume, Airway Flow and Pressure, Acta Anaesthesiol Scand
  • Kestner, J. (1981) The Mechanical Ventilator, In: Rattenborg, C.C.,(Ed.), (1981) Clinical Use of Mechanical Ventilation, YearBook Medical Publishers
  • Mapleson, W.W., (1962) The Effect of Lung Characteristics on the Functioning of Artificial Ventilators, Anaesthesia
  • Pilbeam, S.P., (1986), Mechanical Ventilation: Physiological and Clinical Applications, MultiMedia Publishing
  • Rendell-Baker, L., Pettis, J.L., (1987), The Development of Positive Pressure Ventilators, In: Atkinson, R.S., Boulton, T.B., (Eds.), (1987), The History of Anaesthesia, Royal Society of Medicine
  • Rowland, M. (1992), Biology, Nelson Publishing
  • Young, J.D., Sykes, M.K., (1994) Artificial Ventilation: history, equipment and techniques, In: Moxham, J., Goldstone, J., (Eds.), (1994) Assisted Ventilation 2nd Edition, BMJ Publishing Group

Engineering Essays




order personalized engineering essay today


No Plagiarism Guarantee



Fully confidential Service



3 Hour and Next Day Rush Service



Delivered on Time or Free



Free Plagiarism Report with Every Essay Order



Your essay will never be resold



7 Days for Amendment Requests



1st Class or 2:1 standard guaranteed



All essays written to exact specifications



All Essays are Fully Referenced



100% Complete Satisfaction Guaranteed

Custom essays | Free coursework essays | Our guarantees | Our essay prices | Essay writing tips | Vacancies for essay writers | FAQs

Sister sites: Law Articles | Term Papers | Essays | Law Essays | English Literature Essays

© 2008 Academic Answers Limited | Get Verified | Custom Essays and Free Coursework | RSS | Sitemap

Safe Purchasing Guarantee

A UK Based Company Registered in England and Wales - Registration No: 4964706 - VAT Registration No: 842417633